VITALITY 2
Report
- Report Number
- 2124215-2011-00999
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- January 13, 2011
- Report Date
- February 11, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
AS OF TODAY, THERE IS NO ADDITIONAL INFORMATION AVAILABLE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.
THE DEVICE IS CURRENTLY UNDERGOING ANALYSIS. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.
SUBSEQUENT INFORMATION INDICATED THAT THE DEVICE WAS EXPLANTED AND HAS BEEN RETURNED FOR ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION VIA A LATITUDE RED ALERT THAT THIS DEVICE DECLARED A BATTERY STATUS OF END OF LIFE (EOL). AS OF TODAY, NO INTERVENTION HAS BEEN PERFORMED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. THE DEVICE REMAINS IN SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention | 0148| T177| 1852 |