FDA Adverse Event Injury Summary report: N

VITALITY 2

MDR report key: 2041643 · Received April 6, 2011

Report

Report Number
2124215-2011-01390
Event Type
Injury
Date Received
April 6, 2011
Date of Event
January 14, 2011
Report Date
March 17, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED FOR ANALYSIS. A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE DEVICE DECLARED END OF LIFE (EOL) POST EXPLANT AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Additional Manufacturer Narrative · 1

THIS DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, OR IF THE DEVICE IS RETURNED, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) DUE TO EXTENDED CHARGE TIMES EXCEEDING THE EXTENDED CHARGE TIME LIMIT. THE DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T165

Patients

Seq Age Sex Outcome Treatment
1 83 YR Required Intervention 0154| T165| E110| 1861| 4473