Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED A DISLODGED DISPLAY SCREEN, PARTIALLY DISLODGED FORCE SENSOR PINS, AND OUT OF CALIBRATION FORCE SENSOR. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED, THIS FAILURE WAS DUPLICATED DURING INVESTIGATION.
Description of Event or Problem · 1
PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED PARTIALLY DISLODGED FORCE SENSOR PINS AND FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |