FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2041636 · Received March 10, 2011

Report

Report Number
2531779-2011-01461
Event Type
Malfunction
Date Received
March 10, 2011
Report Date
February 10, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH WAS NOT DUPLICATED DURING INVESTIGATION.

Description of Event or Problem · 1

THE PUMP WAS RETURNED TO ANIMAS DUE TO LOSS OF PRIME ALARMS. THE PUMP WAS EVALUATED AND THE FORCE SENSOR PINS WERE FOUND TO BE PARTIALLY DISLODGED. THIS REPORT IS BEING MADE BASED ON THE EVAL RESULTS.