FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2041586
·
Received March 10, 2011
Report
- Report Number
- 2531779-2011-01490
- Event Type
- Malfunction
- Date Received
- March 10, 2011
- Report Date
- February 10, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED, WHICH COULD NOT BE DUPLICATED DURING TESTING.
Description of Event or Problem · 1
PUMP RETURNED TO ANIMAS DUE TO LOSS OF PRIME ALARMS. THE PUMP WAS EVALUATED AND THE FORCE SENSOR PINS WERE FOUND TO BE PARTIALLY DISLODGED. THIS REPORT IS BEING MADE BASED ON THE EVAL RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |