FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2041523
·
Received March 24, 2011
Report
- Report Number
- 1720753-2011-02761
- Event Type
- Malfunction
- Date Received
- March 24, 2011
- Date of Event
- March 8, 2011
- Report Date
- March 24, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE FLUORO FUNCTIONS BOARD, ADJUSTED THE 5V POWER SUPPLY, AND RELOADED CALIBRATION FILES. THE SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM LOCKED UP DURING A PROCEDURE. NO PATIENT INJURY WAS REPORTED.