FDA Adverse Event Injury Summary report: N

FLEXTEND II

MDR report key: 2041499 · Received April 6, 2011

Report

Report Number
2124215-2011-00854
Event Type
Injury
Date Received
April 6, 2011
Date of Event
January 12, 2011
Report Date
January 12, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS THIS LEAD WILL NOT BE RETURNED NO ANALYSIS WILL BE CONDUCTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS EVENT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD WAS EXPLANTED DUE TO INCREASED THRESHOLDS, DECREASED SENSING, INCREASED IMPEDANCES AND AN X-RAY CONFIRMED LEAD DAMAGE IN THE AREA OF THE SUBCLAVIAN. A NEW LEAD WAS IMPLANTED. THIS RA LEAD WAS CAPPED/ABANDONED AND WILL NOT BE RETURNED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXTEND II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4096

Patients

Seq Age Sex Outcome Treatment
1 H197| 1852| 4096| 0148