FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2041494
·
Received March 10, 2011
Report
- Report Number
- 3002158293-2011-00272
- Event Type
- Malfunction
- Date Received
- March 10, 2011
- Date of Event
- February 3, 2011
- Report Date
- March 7, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHARGER NOT FUNCTIONING PROPERLY) WAS CONFIRMED. UPON EVALUATION, THE POWER BRICK CONNECTOR WAS FOUND TO BE DAMAGED AND WOULD NOT STAY CONNECTED TO THE CHARGER. THE ROOT CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY DETERMINED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE POWER BRICK. THE PATIENT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.
Description of Event or Problem · 1
A (B)(6) MALE PATIENT'S WIFE CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT'S BATTERY CHARGER/MODEM WOULD ONLY WORK IF THE CORD WAS HELD IN PLACE. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |