FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 2041493 · Received April 6, 2011

Report

Report Number
2124215-2011-00830
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
January 13, 2011
Report Date
January 13, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INITIAL INTERROGATION REVEALED THAT THE DEVICE HAD REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). OUR TECHNICIANS COMPARED THE OBSERVED RATE OF BATTERY USAGE TO THE EXPECTED RATE OF BATTERY USAGE. THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE AND PROGRAMMED SETTINGS. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, TWO CONSECUTIVE CHARGE TIMES IN EXCESS OF THE 18 SECOND SPECIFICATION TRIGGERED ERI EARLIER THAN EXPECTED. ENGINEERS CONCLUDED THAT THIS DEVICE DID NOT EXPERIENCE A COMPONENT FAILURE OR PREMATURE BATTERY DEPLETION. RATHER, THE ERI CHARGE TIME INDICATOR WAS DECLARED DUE TO A BUILD-UP OF INTERNAL BATTERY IMPEDANCE, PREVENTING THE DEVICE FROM MEETING ITS EXPECTED LONGEVITY PER DEVICE LABELING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED THIS DEVICE AT THEIR POST MARKET QUALITY ASSURANCE LABORATORY FOR A ROUTINE ASSESSMENT POST EXPLANT. NO PERFORMANCE OR LONGEVITY ALLEGATIONS COMMUNICATED WITH THE RETURNED PRODUCT. ADDITIONALLY, NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H155

Patients

Seq Age Sex Outcome Treatment
1 68 YR H155| 4096| 0185| N119