FDA Adverse Event Malfunction Summary report: N

VITALITY

MDR report key: 2041396 · Received April 6, 2011

Report

Report Number
2124215-2011-00852
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
January 13, 2011
Report Date
January 25, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Additional Manufacturer Narrative · 1

ANALYSIS IS CURRENTLY ON-GOING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR WAS EXPLANTED AND REPLACED FOR AN UNSPECIFIED REASON. NO FIELD ALLEGATIONS HAVE BEEN RECEIVED AGAINST THIS PRODUCT. THIS REPORT WAS CREATED DUE TO LABORATORY FINDINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T127

Patients

Seq Age Sex Outcome Treatment
1 53 YR 1851| (B)(4)| 0127| T127| E110