Report
- Report Number
- 1720753-2011-02697
- Event Type
- Malfunction
- Date Received
- March 23, 2011
- Date of Event
- March 4, 2011
- Report Date
- March 23, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE COLLIMATOR WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9900 SYSTEM HAD A COLLIMATOR IRIS POTENTIOMETER ERROR AT BOOT UP. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |