FDA Adverse Event Malfunction Summary report: N

CONTOUR CURVED CUTTER STAPLER

MDR report key: 2041338 · Received April 6, 2011

Report

Report Number
3005075853-2011-01360
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 14, 2011
Report Date
March 15, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K040038
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, ALTHOUGH THE STAPLES OF THE ORAL SIDE WERE FORMED PROPERLY, THE STAPLES OF THE OTHER SIDE WERE UNFORMED WHEN THE DEVICE WAS FIRED ON THE RECTUM AT THE FIRST FIRING. THE UNFORMED STAPLE LINE WAS SUTURED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. REINFORCEMENT MATERIAL WAS NOT USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR CURVED CUTTER STAPLER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA G4TR7M

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE