CONTOUR CURVED CUTTER STAPLER
Report
- Report Number
- 3005075853-2011-01360
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- March 14, 2011
- Report Date
- March 15, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K040038
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS SHOWED THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITIONS AND WITH A CARTRIDGE RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED VOID OF STAPLES, WITH THE WASHER COMPLETELY CUT, WITH THE DRIVERS EXPOSED AND WITH THE KNIFE RECESS BELOW THE CARTRIDGE DECK. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH AN ENGINEERING SAMPLE RELOAD AND THE DEVICE FIRED, FORMING ALL STAPLES AND CUTTING AS INTENDED. THE CUT LINE WAS COMPLETE, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE DEVICE LOCKOUT AND THE RELOAD LOCKOUT WERE FUNCTIONAL. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, ALTHOUGH THE STAPLES OF THE ORAL SIDE WERE FORMED PROPERLY, THE STAPLES OF THE OTHER SIDE WERE UNFORMED WHEN THE DEVICE WAS FIRED ON THE RECTUM AT THE FIRST FIRING. THE UNFORMED STAPLE LINE WAS SUTURED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. REINFORCEMENT MATERIAL WAS NOT USED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR CURVED CUTTER STAPLER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | G4TR7M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARTRIDGE |