FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2041328 · Received March 23, 2011

Report

Report Number
1720753-2011-02684
Event Type
Malfunction
Date Received
March 23, 2011
Date of Event
March 2, 2011
Report Date
March 23, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE 12 VOLT POWER SUPPLY WAS ADJUSTED AND THE GENERATOR AND FILAMENT WERE CALIBRATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM HAD A LOW MILLIAMPS ERROR AND THE MONITOR WOULD NOT BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1