FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2041317
·
Received March 23, 2011
Report
- Report Number
- 1720753-2011-02671
- Event Type
- Malfunction
- Date Received
- March 23, 2011
- Date of Event
- March 4, 2011
- Report Date
- March 23, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE MANUFACTURER'S SERVICE REPRESENTATIVE PERFORMED AN ON-SITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE HIGH VOLTAGE CABLE AND CALIBRATED THE SYSTEM. THE SYSTEM WAS TESTED AND IS OPERATING AS INTENDED. THIS EVENT MAY BE AN ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE 9800 SYSTEM DISPLAYED A COLLIMATOR IRIS ERROR MESSAGE. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |