FDA Adverse Event Malfunction Summary report: N

BD EPICENTER¿ SINGLE USER SOFTWARE

MDR report key: 20412164 · Received October 9, 2024

Report

Report Number
1119779-2024-00746
Event Type
Malfunction
Date Received
October 9, 2024
Date of Event
September 10, 2024
Report Date
January 23, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JQP
UDI-DI
00382904439729
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: CUSTOMER REPORTING AN ISSUE IS REGARDING THE PUD VERSION 7.41 WHICH HAS THE RESISTANCE BREAKPOINTS FOR S. AUREUS AND OTHER COAG.NEGATIVE STAPHYLOCOCCI >/=2, WHILST ON THE EUCAST2024 DOCUMENT IS R>2. THIS MEANS THAT IN THE EPICENTER (443972/G6MY646007FV) BREAKPOINT TABLE IN PUD 7.41 WE SHOULD HAVE R >/=4. IN THE EUCAST 2023 DOCUMENT, THE CIP BREAKPOINT FOR STAPH AUREUS AND STAPH COAG- WAS NO SUSCEPTIBLE BREAKPOINT AND A RESISTANT OF >1 (IN THE PUD WE SET THE S <= 1 AND THE R >= 2 WITH AN ASSOCIATED RULE (985) THAT CHANGES ANY RESULT THAT IS AN S (SUSCEPTIBLE) TO I (INTERMEDIATE). IN THE EUCAST 2024 DOCUMENT, THE CIP BREAKPOINT FOR STAPH AUREUS AND STAPH COAG- WAS UPDATED TO NO SUSCEPTIBLE OR INTERMEDIATE BREAKPOINT AND A RESISTANT OF >2. THE PUD V7.41 RELEASE NOTES STATE THAT THE BREAKPOINT WAS NOT UPDATED DUE TO A PENDING REVIEW. THE CUSTOMER IS ABLE TO CUSTOMIZE THE BREAKPOINT IF DESIRED AND UPDATE PER THE EUCAST DOCUMENT. THIS IS NOT A CONFIRMED COMPLAINT OF A BD PRODUCT. COMPLAINTS FOR SOFTWARE WERE UNDER STATISTICAL CONTROL FOR THE MONTH OF SEPTEMBER. NO TRENDS INDICATED. DEVICE HISTORY RECORD REVIEW IS NOT REQUIRED FOR STANDALONE SOFTWARE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED.

Additional Manufacturer Narrative · 0

E.1. INITIAL REPORTER FACILITY NAME: (B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD EPICENTER¿ SINGLE USER SOFTWARE THERE WAS ONE FALSE RESULT DUE TO A MISTAKEN BREAKPOINT AFTER THE PUD UPDATE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD EPICENTER¿ SINGLE USER SOFTWARE THERE WAS ONE FALSE RESULT DUE TO A MISTAKEN BREAKPOINT AFTER THE PUD UPDATE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1790470 BD EPICENTER¿ SINGLE USER SOFTWARE CALCULATOR/DATA PROCESSING MODULE JQP BECTON DICKINSON & CO. (SPARKS) UNKNOWN 00382904439729

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown