FDA Adverse Event Injury Summary report: N

UNKNOWN ZIMMER FEMORAL STEM

MDR report key: 2041110 · Received March 29, 2011

Report

Report Number
1822565-2011-00783
Event Type
Injury
Date Received
March 29, 2011
Date of Event
October 30, 2007
Report Date
March 1, 2011
Manufacturer
ZIMMER, INC.
Product Code
KWY
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE STERILIZATION PROCESS FOR ALL DEVICES PRODUCED BY ZIMMER ARE VALIDATED IN ACCORDANCE WITH FDA REGULATIONS AND ISO STANDARDS TO A STERILITY ASSURANCE LEVEL (SAL) OF 1.0 X 10 (-6) OR BETTER, AND ARE PROCESSED ACCORDING TO THE VALIDATED STERILIZATION PROCESS PARAMETERS AND MUST MEET ALL THE ACCEPTANCE CRITERIA FOR STERILITY RELEASE. THEREFORE, IT IS HIGHLY UNLIKELY THAT THE SPECIFIED DEVICE CAUSED OR CONTRIBUTED TO ANY PT INFECTION. IN ADDITION, BEFORE EACH MFG LOT IS RELEASED, THE "CERTIFICATE OF PROCESSING" FROM THE STERILIZATION SUPPLIER IS REVIEWED FOR CONFORMANCE. THIS CERTIFICATE LISTS THE MAXIMUM, MINIMUM AND ACTUAL GAMMA DOSE IN KGY. WITH THE INFO PROVIDED, A DEFINITIVE ROOT CAUSE OF THE ALLEGED CONDITION CANNOT BE DETERMINED. REVIEW OF THE DEVICE HISTORY RECORDS WAS NOT POSSIBLE AS THE PRODUCT AND/OR LOT NUMBERS REQUIRED FOR RETRIEVAL WERE UNAVAILABLE. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE AVAILABLE INFO, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PT WAS TREATED FOR A LARGE HEMATOMA AND INFECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNKNOWN ZIMMER FEMORAL STEM HIP PROSTHESIS KWY ZIMMER, INC.

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention