VITALITY
Report
- Report Number
- 2124215-2011-00820
- Event Type
- Injury
- Date Received
- April 6, 2011
- Date of Event
- January 11, 2011
- Report Date
- January 14, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.
AT THIS TIME THE DEVICE HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC, (B)(4) FOR ANALYSIS. THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF FURTHER INFORMATION BECOMES AVAILABLE THIS EVENT WILL BE REOPENED.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DECLARED END OF LIFE (EOL) DUE TO A CHARGE TIME MEASUREMENT OF 30.0 SECONDS. THE MONITORING VOLTAGE WAS 2.65 VOLTS. THE DEVICE WAS EXPLANTED AND SUCCESSFULLY REPLACED. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Required Intervention | 0185| T180 |