FDA Adverse Event Malfunction Summary report: N

7900

MDR report key: 2041022 · Received March 23, 2011

Report

Report Number
9680959-2011-00857
Event Type
Malfunction
Date Received
March 23, 2011
Date of Event
March 7, 2011
Report Date
March 23, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE FAULTY PART WAS RECONFIGURED AND THUS REPAIRED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO PRODUCE FLUOROSCOPY X-RAY. NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 7900 FLUOROSCOPIC XRAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1