FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 2040828 · Received April 6, 2011

Report

Report Number
2124215-2011-00752
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
July 25, 2010
Report Date
January 11, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT WITH THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) INQUIRED IF THE DEVICE WOULD SET OFF ALARMS IN DEPARTMENT STORES. THE DEVICE WAS SUBSEQUENTLY EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THE DEVICE WAS RETURNED FOR DISPOSAL AFTER EXPLANT. ROUTINE INITIAL RETURNED DEVICE TESTING INDICATED THAT DETAILED ANALYSIS WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H177

Patients

Seq Age Sex Outcome Treatment
1 59 YR 0184| T177| H177