CONTAK RENEWAL
Report
- Report Number
- 2124215-2011-00752
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- July 25, 2010
- Report Date
- January 11, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT WITH THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) INQUIRED IF THE DEVICE WOULD SET OFF ALARMS IN DEPARTMENT STORES. THE DEVICE WAS SUBSEQUENTLY EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THE DEVICE WAS RETURNED FOR DISPOSAL AFTER EXPLANT. ROUTINE INITIAL RETURNED DEVICE TESTING INDICATED THAT DETAILED ANALYSIS WAS REQUIRED.