FDA Adverse Event Malfunction Summary report: N

VITALITY

MDR report key: 2040795 · Received April 6, 2011

Report

Report Number
2124215-2011-01152
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
January 10, 2011
Report Date
January 10, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS SUCCESSFULLY INTERROGATED BY BOSTON SCIENTIFIC'S POST MARKET QUALITY ASSURANCE LABORATORY. ENGINEERING CALCULATIONS DETERMINED THAT THIS DEVICE DID NOT MEET EXPECTED LONGEVITY. PACING, SENSING, AND SHOCKING FUNCTIONS WERE VERIFIED. THE RECORDED PACING AND SHOCK IMPEDANCE MEASUREMENTS WERE NORMAL. A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION FROM AN IMPLANT FORM THAT THIS DEVICE WAS ELECTIVELY EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THERE WERE NO ALLEGATIONS OR COMPLAINTS AGAINST THE DEVICE'S OPERATION, FUNCTIONALITY OPR LONGEVITY. THE DEVIE WAS RECEIVED AT BOSTON SCIENTIFIC'S RETURN PRODUCTS DEPARTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T135

Patients

Seq Age Sex Outcome Treatment
1 63 YR E102| T135| 0174