VITALITY
Report
- Report Number
- 2124215-2011-00834
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- January 10, 2011
- Report Date
- January 10, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RETURN TO BOSTON SCIENTIFIC'S POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THE DEVICE DECLARED ITS ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. ADDITIONAL LAB ANALYSIS AND TESTING SHOWED ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF DEVICE INTERNAL BATTERY IMPEDANCE. THE EOL (END OF LIFE) INDICATOR WAS NOT DECLARED WHILE THE DEVICE WAS IN SERVICE. TESTING CONFIRMED ALL DEVICE THERAPIES WERE AVAILABLE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. THERE WERE NO KNOWN ALLEGATIONS AGAINST THE DEVICE, AND NO REPORTS OF ADVERSE PATIENT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T135 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | T135| 0185 |