VITALITY 2
Report
- Report Number
- 2124215-2011-00835
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- January 11, 2011
- Report Date
- February 24, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
ALL AVAILABLE INFORMATION INDICATES THAT THIS PRODUCT REMAINS IN SERVICE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
AT THIS TIME THIS PRODUCT HAS NOT BEEN RETURNED TO BOSTON SCIENTIFIC FOR ANALYSIS. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.
THE DEVICE WAS LATER EXPLANTED WITH AN ALLEGATION OF PREMATURE ERI.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) DUE TO A CHARGE TIME (CT) OF 29.7 SECONDS. TECHNICAL SERVICES (TS) DISCUSSED A POSSIBLE CT ANOMALY WITH THE DEVICE, AND RECOMMENDED REPLACEMENT. IT WAS NOTED THAT THE PATIENT WOULD BE SCHEDULED FOR REPLACEMENT AND THE DEVICE WOULD BE RETURNED FOR ANALYSIS FOLLOWING EXPLANT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
THE DEVICE WAS LATER RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Required Intervention | 0157| T177 |