VITALITY
Report
- Report Number
- 2124215-2011-00524
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Date of Event
- January 7, 2011
- Report Date
- January 25, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- PHYSICIAN
Narratives
A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.
THIS PRODUCT IS CURRENTLY UNDERGOING DETAILED ANALYSIS. THIS EVENT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED FOR UNKNOWN REASONS AND WAS RETURNED FOR DISPOSAL. ROUTINE INITIAL DEVICE TESTING INDICATED THAT DETAILED ANALYSIS WAS REQUIRED. TO DATE, THERE HAVE BEEN NO REPORTED ADVERSE PATIENT EFFECTS RELATED TO THIS CLINICAL OBSERVATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | 0144| T135| 1790| T180 |