FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2040599 · Received March 22, 2011

Report

Report Number
1720753-2011-02624
Event Type
Malfunction
Date Received
March 22, 2011
Date of Event
February 27, 2011
Report Date
March 22, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN OVER THE PHONE INVESTIGATION. THE CIRCUIT BREAKER WAS RESET DURING THE SERVICE CALL. THE SYSTEM WAS FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE 9900 SYSTEM WOULD NOT TURN ON. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1