FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2040586
·
Received March 22, 2011
Report
- Report Number
- 1720753-2011-02602
- Event Type
- Malfunction
- Date Received
- March 22, 2011
- Date of Event
- March 1, 2011
- Report Date
- March 22, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE LIFT MOTOR POWER SUPPLY WAS REPLACED AND THE COLLIMATOR WAS ADJUSTED. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE UNIT WOULD NOT MOVE UP AND DOWN AND IRIS LEAVES APPEAR AT THE BEGINNING. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|