GAS SUPPLY LINE 2.1MT
Report
- Report Number
- 9611451-2011-00217
- Event Type
- Malfunction
- Date Received
- April 6, 2011
- Report Date
- March 7, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). WE ARE CURRENTLY ENDEAVOURING TO OBTAIN FURTHER INFORMATION WITH REGARD TO THIS COMPLAINT. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE TWO RETURNED COMPLAINT BC400 GAS SUPPLY LINES WERE OCCLUDED AT ONE CONNECTOR END AND SUBMERGED IN A WATERBATH TO TEST FOR LEAKS. THE WATERBATH TEST WAS REPEATED BY OCCLUDING THE OTHER CONNECTOR END. RESULTS: THE WATERBATH TEST DID NOT REVEAL ANY LEAKAGE FOR BOTH RETURNED COMPLAINT SUPPLY LINES. NO LOT CHECK COULD BE PERFORMED AS NO LOT NUMBER WAS PROVIDED. CONCLUSION: NO FAULT WAS FOUND WITH TWO RETURNED COMPLAINT GAS SUPPLY LINES.
A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS AIR LEAKING AROUND THE ADAPTOR OF AN BC400 GAS SUPPLY LINE DURING USE. NO PATIENT CONSEQUENCE WAS REPORTED.
A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS AIR LEAKING AROUND THE ADAPTOR OF AN BC400 GAS SUPPLY LINE DURING USE.NO PATIENT CONSEQUENCE WAS REPORTED.