FDA Adverse Event Malfunction Summary report: N

GAS SUPPLY LINE 2.1MT

MDR report key: 2040493 · Received April 6, 2011

Report

Report Number
9611451-2011-00217
Event Type
Malfunction
Date Received
April 6, 2011
Report Date
March 7, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). WE ARE CURRENTLY ENDEAVOURING TO OBTAIN FURTHER INFORMATION WITH REGARD TO THIS COMPLAINT. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE TWO RETURNED COMPLAINT BC400 GAS SUPPLY LINES WERE OCCLUDED AT ONE CONNECTOR END AND SUBMERGED IN A WATERBATH TO TEST FOR LEAKS. THE WATERBATH TEST WAS REPEATED BY OCCLUDING THE OTHER CONNECTOR END. RESULTS: THE WATERBATH TEST DID NOT REVEAL ANY LEAKAGE FOR BOTH RETURNED COMPLAINT SUPPLY LINES. NO LOT CHECK COULD BE PERFORMED AS NO LOT NUMBER WAS PROVIDED. CONCLUSION: NO FAULT WAS FOUND WITH TWO RETURNED COMPLAINT GAS SUPPLY LINES.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS AIR LEAKING AROUND THE ADAPTOR OF AN BC400 GAS SUPPLY LINE DURING USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THERE WAS AIR LEAKING AROUND THE ADAPTOR OF AN BC400 GAS SUPPLY LINE DURING USE.NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GAS SUPPLY LINE 2.1MT BZO BZO FISHER & PAYKEL HEALTHCARE LTD BC400-10