ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2011-00216
- Event Type
- Malfunction
- Date Received
- April 5, 2011
- Date of Event
- February 22, 2011
- Report Date
- March 9, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BZE
- PMA / PMN Number
- REFER TO H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NURSE
Narratives
(B)(4). PMA/510(K): THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT IS NOT SOLD IN THE USA, BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. MANUFACTURER NARRATIVE: METHOD: ONLY ONE EVAQUA EXPIRATORY TUBE WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) NEW ZEALAND FOR INSPECTION. IT WAS VISUALLY INSPECTED FOR DAMAGE AND PRESSURE TESTED FOR LEAKS. RESULTS: THE VISUAL INSPECTION REVEALED A HOLE IN THE RETURNED EVAQUA TUBE. THE TUBE FAILED THE PRESSURE TEST AS IT WAS LEAKING FROM THE HOLE. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 100819. CONCLUSION: BASED ON OUR EVALUATION, IT IS LIKELY THAT THE EVAQUA EXPIRATORY TUBE HAS BEEN PULLED, CREATING A HOLE AND RESULTING IN THE REPORTED LEAK. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO DISTRIBUTION. ANY BREATHING CIRCUIT WHICH DOES NOT PASS THE PRESSURE TEST IS DISCARDED. THIS SUGGESTS THAT THE BREATHING CIRCUIT WAS PUNCTURED POST-PRODUCTION. THE KEY DIFFERENCE BETWEEN FISHER & PAYKEL HEALTHCARE'S EVAQUA BREATHING CIRCUITS AND THE CONVENTIONAL BREATHING CIRCUITS IS THAT THE EXPIRATORY TUBE OF EVAQUA CIRCUITS SUCH AS THE RT340 IS COMPOSED OF A THIN, SEMI-PERMEABLE FILM SPECIALLY DESIGNED TO ALLOW WATER VAPOR FROM EXPIRED VENTILATORY GASES TO PASS THROUGH. THE EVAQUA EXPIRATORY TUBE HAS A PROTECTIVE MESH WHICH PREVENTS DAMAGE TO THE WALLS OF THE TUBE, HOWEVER THE EVAQUA TUBING REMAINS MORE SUSCEPTIBLE TO DAMAGE THAN CONVENTIONAL CIRCUITS WHEN EXPOSED TO ROUGH HANDLING OR DAMAGE CAUSED BY OBJECTS SUCH AS CIRCUIT HANGERS. THE RT340 USER INSTRUCTIONS INCLUDE THE FOLLOWING STATEMENT: "SET APPROPRIATE VENTILATOR ALARMS", AS VENTILATOR ALARMS ALERT THE USER TO A LEAK IN THE SYSTEM AND ALLOWS CAREGIVERS TO ACT BY REPLACING THE BREATHING CIRCUIT ACCORDING TO THEIR STANDARD PROCEDURES. IT ALSO ADVISE THE USER TO "FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE". OUR MONITORING AND TRENDING OF COMPLAINTS INVOLVING ADULT EVAQUA BREATHING CIRCUIT LEAKS HAS A RATE OF OCCURRENCE OF 78 DEVICES PER MILLION SOLD IN THE LAST YEAR TO THE END OF (B)(6) 2011.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THE EXPIRATORY TUBES OF TWO RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUITS WERE FOUND LEAKING. THIS WAS NOTICED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LIMITED | RT340 | 100819 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |