FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 20404810 · Received October 9, 2024

Report

Report Number
2955842-2024-20453
Event Type
Malfunction
Date Received
October 9, 2024
Date of Event
September 17, 2024
Report Date
September 17, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874119785
PMA / PMN Number
K214095
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE PROGRASP FORCEPS INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A FRAYED GRIP CABLE AT THE FORCE AMPLIFIER. SOME BUNDLES OF THE CABLE WERE FRAYED, BUT THE CABLE WAS NOT FULLY BROKEN. THE FRAYED CABLE STRANDS STUCK OUT AT THE WRIST. THERE WAS NO EVIDENCE OF DISCOLORATION OR CORROSION/CONTAMINATION ON THE CABLES TO INDICATE A REPROCESSING INDUCED ISSUE. THE COMPONENTS SURROUNDING THE FRAYED CABLE DID NOT EXHIBIT ABNORMAL SHARP EDGES OR DAMAGE THAT WOULD HAVE CONTRIBUTED TO THE CABLE FRAYING. THE INSTRUMENT WAS ALSO FOUND TO HAVE THE GRIP AXIS PIN DISLODGED FROM THE GRIPS. THE PIN WAS RETURNED WITH THE INSTRUMENT. THE PIN WAS FOUND STILL ATTACHED TO THE DISTAL END BUT PROTRUDING OUT FROM THE GRIPS. GRIPS AND OTHER COMPONENTS ADJACENT TO THE DISLODGED PIN DO NOT SHOW DAMAGE. THE COMPLAINT REGARDING FRAYED WIRES AT THE HINGE AND A PROTRUDING SCREW BELOW THE INSTRUMENT TIPS WAS CONFIRMED BY FAILURE ANALYSIS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PROGRASP FORCEPS INSTRUMENT HAD FRAYING WIRES AS THE HINGE. ADDITIONALLY, THE INSTRUMENT WAS ALSO NOTED TO HAVE A PROTRUDING SCREW RIGHT BELOW THE INSTRUMENT TIPS. THERE WAS NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
686804 ENDOWRIST PROGRASP FORCEPS NAY INTUITIVE SURGICAL, INC 471093-11 K12240118 0397 00886874119785

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose DA VINCI INSTRUMENTS AND ACCESSORIES