UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-00909
- Event Type
- Malfunction
- Date Received
- April 5, 2011
- Date of Event
- September 18, 2008
- Report Date
- September 18, 2008
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE SAMPLE WAS PLASMA COLLECTED INTO A LITHIUM HEPARIN GEL TUBE AND CENTRIFUGED AT 2400 RPM FOR 10 MINUTES. PER CUSTOMER, ALL TESTING WAS RUN ON THE PRIMARY TUBES. QC ARE RUN DAILY AND CONFIRMED NO PERFORMANCE ISSUES BEEN NOTED. A BCI FIELD SERVICE ENGINEER (FSE) PERFORMED HIGH SENSITIVITY SYSTEM CHECK ACCUTNI VERIFICATION, WHICH PASSED. FSE INSTRUCTED THE CUSTOMER ON PROPER SAMPLE HANDLING PROCEDURES. FSE VERIFIED REPAIR PER ESTABLISHED PROCEDURES. RESULTS MET PUBLISHED PERFORMANCE VERIFICATIONS. A CLEAR ROOT CAUSE CAN NOT BE DETERMINED FROM THE INFORMATION PROVIDED. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF COMPLAINTS WITHIN THE DATE RANGE OF (B)(4) 2011 - (B)(4) 2011 FOR ADDITIONAL REPORTABLE EVENT/OCCURRENCE.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO NON REPRODUCIBLE ELEVATED ACCUTNI RESULT BELOW THE ACUTE MYOCARDIAL INFARCTION (AMI) CUT-OFF, THE RISK STRATIFICATION RANGE GENERATED BY THE UNICEL DXI 800 IMMUNOASSAY ANALYZER. THE RESULT WAS REPORTED OUT OF THE LABORATORY AND QUESTIONED AFTER A SECOND DRAW RESULTS WITHIN THE REFERENCE RANGE. THE SAMPLE WAS REPEATED AND RESULT WITHIN THE REFERENCE RANGE WAS OBTAINED. THE PATIENT WAS ADMITTED TO THE HOSPITAL; HOWEVER NO INVASIVE TREATMENT WERE PERFORMED AS A RESULT OF THIS EVENT. THE CUSTOMER DID NOT REPORT ANY ISSUES OR CONCERNS IN REGARDS TO PATIENT OR USER ATTRIBUTED OR CONNECTED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXI800 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |