FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2040451 · Received April 5, 2011

Report

Report Number
1423500-2011-04137
Event Type
Malfunction
Date Received
April 5, 2011
Date of Event
March 13, 2011
Report Date
March 13, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT WAS NOT CONFIRMED. BASED ON THE INFORMATION GATHERED DURING BAXTER?S INVESTIGATION, THE ROOT CAUSE OF THE SE 2240 WAS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE BECAUSE, THERE WAS AN OPEN CLAMP ON UNUSED SUPPLY LINE. THE HOME PATIENT (HP) STATED THERE WERE OPEN CLAMPS ON UNUSED SUPPLY LINES. THE HP STATED HIS CLAMPS ON THE ORGANIZER WERE STILL OPEN. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. THE LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR A USE/USER ERROR IDENTIFIED IN THIS INCIDENT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP REPORT WILL BE SUBMITTED UPON THE COMPLETION OF BAXTER'S QUALITY REVIEW.

Description of Event or Problem · 1

THE HOME PATIENT (HP) CONTACTED GLOBAL TECHNICAL SERVICES (GTS) REGARDING A SYSTEM ERROR (SE) 2240 ALARM WHICH OCCURRED ON THE HOMECHOICE (HC) DURING INITIAL DRAIN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP WITH TROUBLESHOOTING. THE HP STATED THERE WERE OPEN CLAMPS ON UNUSED SUPPLY LINES. THE HP STATED HIS CLAMPS ON THE ORGANIZER WERE STILL OPEN. THE TSR EXPLAINED SE 2240 INDICATES AIR ENTERED THE SET UP. THE TSR FURTHER ASSISTED THE HP TO CYCLE POWER TO CLEAR ALARM. THE HP CONFIRMED TO START OVER WITH ALL NEW SUPPLIES. THE TSR REVIEWED PROPER PROCEDURES PER THE USER MANUAL WITH THE HP. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 28 YR HOMECHOICE