FDA Adverse Event Malfunction Summary report: N

LIH

MDR report key: 2040447 · Received April 5, 2011

Report

Report Number
2050012-2011-00881
Event Type
Malfunction
Date Received
April 5, 2011
Date of Event
February 1, 2011
Report Date
February 1, 2011
Manufacturer
BECKMAN COULTER BIOMEDICAL LTD
Product Code
JJY
PMA / PMN Number
CLASS1EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE IS SERUM. NO ADDITIONAL INFORMATION IS AVAILABLE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT SERUM INDICES (LIH - LIPAEMIA, ICTERUS, HAEMOLYSIS) WERE RUN ON (B)(6) 2010 ON A SAMPLE FROM A PATIENT WITH ACUTE LIVER FAILURE OF UNDEFINED AETIOLOGY. ON (B)(6) 2011, ALTHOUGH THE SAMPLE WAS ICTERIC AND HAD A TOTAL BILIRUBIN (TBIL) CONCENTRATION OF > 50 MG/DL, ALL THREE INDICES GAVE NORMAL RESULTS. A SECOND SAMPLE COLLECTED 12 HOURS LATER GAVE THE EXPECTED RESULTS WHERE ALL 3 INDICES SHOWED ABN INDICATING THAT THE SAMPLE HAD ABNORMAL OPTICAL CHARACTERISTICS AND REQUIRED VISUAL INSPECTION. LIVER TRANSPLANTATION WAS PERFORMED ON (B)(6) 2011 TBIL REMAINED ELEVATED FOR SEVERAL DAYS THEREAFTER AND THE SERUM INDICES CONTINUED TO SHOW MISLEADING RESULTS UP UNTIL (B)(6) 2011. FROM (B)(6) 2011, THE INDICES PERFORMED AS EXPECTED. PATIENT IMPACT IS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIH SERUM INDICES (LIH, LIPAEMIA, ICTERUS, HAEMOLYSIS) JJY BECKMAN COULTER BIOMEDICAL LTD OSR62166 N/A

Patients

Seq Age Sex Outcome Treatment
1