FDA Adverse Event Malfunction Summary report: N

60" HIGH FLOW RATE EXTENSION SET

MDR report key: 2040415 · Received April 5, 2011

Report

Report Number
6000001-2011-02585
Event Type
Malfunction
Date Received
April 5, 2011
Date of Event
March 1, 2011
Report Date
March 9, 2011
Manufacturer
BAXTER HEALTHCARE - AIBONITO
Product Code
FPA
PMA / PMN Number
K811078
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION; HOWEVER, IT HAS NOT YET BEEN COMPLETED. ONCE THE EVALUATION IS COMPLETE, A FOLLOW UP REPORT WILL BE SUBMITTED. A BATCH REVIEW COULD NOT BE PERFORMED AS THE LOT NUMBER IS UNKNOWN.

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER RETURNED THE ACTUAL SAMPLE FOR EVALUATION. THE SAMPLE WAS VISUALLY INSPECTED AND THE REPORTED CONDITION WAS CONFIRMED. THE MALE LUER WAS OBSERVED TO BE BROKEN. THIS COMPONENT IS PURCHASED FROM A SUPPLIER AND A SUPPLIER CORRECTIVE ACTION REPORT WAS SUBMITTED TO THEM. AN ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. A BATCH REVIEW COULD NOT BE PERFORMED AS THE LOT NUMBER IS UNKNOWN. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Description of Event or Problem · 1

THE CUSTOMER RETURNED AN ADDITIONAL SAMPLE OF THE 60" HIGH FLOW RATE EXTENSION SET, PRODUCT CODE 2N3349, LOT NUMBER UNKNOWN, FOR EVALUATION. THROUGH EVALUATION, THE RETURNED SAMPLE WAS DETERMINED TO BE BROKEN. THE MALE LUER TIP WAS RECEIVED BROKEN. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITION INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 60" HIGH FLOW RATE EXTENSION SET SET, ADMINISTRATION, INTRAVASCULAR FPA BAXTER HEALTHCARE - AIBONITO

Patients

Seq Age Sex Outcome Treatment
1