FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2040388
·
Received February 15, 2011
Report
- Report Number
- 1720753-2011-02256
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- February 23, 2011
- Report Date
- March 15, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE MANUFACTURER'S SERVICE REPRESENTATIVE PERFORMED AN ON-SITE INVESTIGATION. THE CINE DRIVE WAS RE-SEATED. THE SYSTEM CALIBRATION FILES WERE RE-LOADED. THE SYSTEM WAS TESTED AND IS OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A NO CINE AVAILABLE ERROR MESSAGE DISPLAYED ON THE 9800 SYSTEM. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |