FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 20402868 · Received October 8, 2024

Report

Report Number
1645337-2024-11775
Event Type
Injury
Date Received
October 8, 2024
Date of Event
July 23, 2024
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001256
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON NOVEMBER 08, 2024, MENTOR RECEIVED ADDITIONAL INFORMATION REGARDING THE PATIENT'S REPLACEMENT DEVICE. IT WAS REPORTED THAT THE PATIENT'S REPLACEMENT DEVICE WAS A 325CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST PROSTHESIS (CATALOG NUMBER 3501650) WITH LOT NUMBER 9780467. IT WAS MENTIONED THAT AT TIME OF DEVICE EXPLANTATION, A PIECE OF TISSUES WAS FOUND IN THE PORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. IT IS UNKNOWN AT THIS TIME IF THE DEVICE WILL BE MADE AVAILABLE FOR RETURN. AS A RESULT, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE HAS BEEN RECEIVED. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON OCTOBER 09, 2024, MENTOR RECEIVED ADDITIONAL INFORMATION REPORTING THAT THE DEVICE WAS TO BE EXPLANTED ON (B)(6) 2024. SECTION D6B. EXPLANTATION DATE: (B)(6) 2024. SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: DEFLATION. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON DECEMBER 05, 2024, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. ON DECEMBER 09, 2024, THE EVALUATION FOR THE DEVICE WAS COMPLETED. DEVICE EVALUATION SUMMARY: DURING VISUAL ANALYSIS OF THE RETURNED SAMPLE SAL SMOOTH RND DIAP 325CC NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED. LEAK TESTING WAS PERFORMED, ACCORDING TO MENTOR PROCEDURE, AND IT IDENTIFIED A TEAR ON THE ANTERIOR VIEW MEASURING APPROXIMATELY 0.2 CM. MICROSCOPIC EXAMINATION WAS PERFORMED ON THE EDGES OF THE RUPTURE, AND PARALLEL STRIATIONS WERE FOUND IN THE WHOLE AREA OF THE TEAR. PARALLEL STRIATIONS ARE CONSISTENT WITH MARKINGS MADE BY A SHARP OBJECT PERFORATING THE IMPLANT SHELL. BASED ON THE INFORMATION CURRENTLY AVAILABLE, THE MICROSCOPIC EXAMINATION OF THE RETURNED PRODUCT INDICATES THAT THE IMPLANT COULD HAVE BEEN DAMAGED DURING OR SUBSEQUENT TO IMPLANTATION. THE PRODUCT INSERT DATA SHEET CAUTIONS TO NOT ALLOW CAUTERY DEVICES OR SHARP INSTRUMENTS, SUCH AS SCALPELS, SUTURE NEEDLES, HYPODERMIC NEEDLES, HEMOSTATS, ADSON FORCEPS, OR SCISSORS TO CONTACT THE DEVICE DURING THE IMPLANTATION OR OTHER SURGICAL PROCEDURES. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A 80-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH 325CC MENTOR SMOOTH ROUND MODERATE PROFILE SALINE BREAST PROSTHESIS AND EXPERIENCED A DEFLATION ON THE LEFT SIDE POST-OPERATIVELY, WHICH WAS DIAGNOSED VIA PHOTOS. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21989 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 9966282 00081317001256

Patients

Seq Age Sex Outcome Treatment
1 80 YR Female Other| R