ISITE PACS
Report
- Report Number
- 2954704-2011-00004
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- January 7, 2011
- Report Date
- March 15, 2011
- Manufacturer
- PHILIPS HEALTHCARE INFORMATICS, INC.
- Product Code
- LLZ
- PMA / PMN Number
- K063267
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
ISITE PACS SOFTWARE PACKAGE VER.4X IS USED WITH GENERAL COMPUTING HARDWARE TO ACQUIRE, STORE, DISTRIBUTE, PROCESS AND DISPLAY IMAGES AND ASSOCIATED DATA THROUGHOUT A CLINICAL ENVIRONMENT. THE MALFUNCTION IS CURRENTLY UNDER DEFECT INVESTIGATION AND HEALTH HAZARD EVALUATION. THE DEVICE IS SOFTWARE AND RESIDES AT THE CUSTOMER FACILITY.
THE CUSTOMER IDENTIFIED THE FOLLOWING BEHAVIOR DURING TESTING: PATIENT A = (B)(6), PATIENT B = (B)(6). PATIENT A'S RECORDS ARE LINKED BETWEEN MULTIPLE FACILITIES (LOCATIONS). PATIENT B IS IDENTIFIED AS THE SAME PERSON AS PATIENT A, AND A MERGE IS CONDUCTED BETWEEN A AND B WITHIN ONE FACILITY. THE LINKS FOR PATIENT A AT ALL FACILITIES WILL BE RETAINED IF PATIENT A IS RETAINED IN THE MERGE AND PATIENT B IS ABSORBED. THE LINKS FOR PATIENT A AT ALL FACILITIES WILL BE LOST IF PATIENT B IS RETAINED IN THE MERGE AND PATIENT A IS ABSORBED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ISITE PACS | ISITE PACS | LLZ | PHILIPS HEALTHCARE INFORMATICS, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |