FDA Adverse Event Malfunction Summary report: N

ISITE PACS

MDR report key: 2040285 · Received February 15, 2011

Report

Report Number
2954704-2011-00004
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
January 7, 2011
Report Date
March 15, 2011
Manufacturer
PHILIPS HEALTHCARE INFORMATICS, INC.
Product Code
LLZ
PMA / PMN Number
K063267
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ISITE PACS SOFTWARE PACKAGE VER.4X IS USED WITH GENERAL COMPUTING HARDWARE TO ACQUIRE, STORE, DISTRIBUTE, PROCESS AND DISPLAY IMAGES AND ASSOCIATED DATA THROUGHOUT A CLINICAL ENVIRONMENT. THE MALFUNCTION IS CURRENTLY UNDER DEFECT INVESTIGATION AND HEALTH HAZARD EVALUATION. THE DEVICE IS SOFTWARE AND RESIDES AT THE CUSTOMER FACILITY.

Description of Event or Problem · 1

THE CUSTOMER IDENTIFIED THE FOLLOWING BEHAVIOR DURING TESTING: PATIENT A = (B)(6), PATIENT B = (B)(6). PATIENT A'S RECORDS ARE LINKED BETWEEN MULTIPLE FACILITIES (LOCATIONS). PATIENT B IS IDENTIFIED AS THE SAME PERSON AS PATIENT A, AND A MERGE IS CONDUCTED BETWEEN A AND B WITHIN ONE FACILITY. THE LINKS FOR PATIENT A AT ALL FACILITIES WILL BE RETAINED IF PATIENT A IS RETAINED IN THE MERGE AND PATIENT B IS ABSORBED. THE LINKS FOR PATIENT A AT ALL FACILITIES WILL BE LOST IF PATIENT B IS RETAINED IN THE MERGE AND PATIENT A IS ABSORBED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ISITE PACS ISITE PACS LLZ PHILIPS HEALTHCARE INFORMATICS, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1