DREAMTOME RX SPHINCTEROTOME
Report
- Report Number
- 3005099803-2011-01164
- Event Type
- Malfunction
- Date Received
- April 5, 2011
- Date of Event
- February 10, 2011
- Report Date
- March 17, 2011
- Manufacturer
- BOSTON SCIENTIFIC - SPENCER
- Product Code
- KNS
- PMA / PMN Number
- K013153
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE WORKING LENGTH WAS TWISTED, THE EXPOSED CUTTING WIRE WAS DISCOLORED, AND THE CUTTING WIRE APPEARED TO HAVE MELTED/SPLIT THE EXTRUSION AT THE DISTAL PIERCE HOLE. THE SPLIT EXTRUSION CAUSED THE CUTTING WIRE ANCHOR TO SLIDE PROXIMALLY FROM THE DISTAL PIERCE HOLE AND ALTERED THE EXPOSED CUTTING WIRE LENGTH TO BECOME SHORTER. A FUNCTIONAL EVALUATION FOUND THAT THE DEVICE DOES NOT MEET THE BOWING SPECIFICATION OF 90 DEGREE MINIMUM DUE TO THE MELTED EXTRUSION (ELONGATED DISTAL PIERCE HOLE) AND THE ALTERED EXPOSED CUTTING WIRE LENGTH. THE CONDITION OF THE RETURNED INCIDENT DEVICE WAS CONSISTENT WITH THE COMPLAINT THAT THE DEVICE FAILED TO BOW COMPLETELY. HOWEVER, THIS CONDITION WAS ATTRIBUTED TO THE MELTED EXTRUSION AND THE ALTERED EXPOSED CUT WIRE LENGTH. DURING MANUFACTURING, THE SPHINCTEROTOME DEVICES ARE 100% INSPECTED AND THE MELTED/SPLIT EXTRUSION IS LIKELY DUE TO PROCEDURAL FACTORS. THEREFORE, THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT. THE DEVICE HISTORY RECORD REVIEW FOUND THAT THE DEVICE MET ALL MANUFACTURING SPECIFICATIONS.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A DREAMTOME RX SPHINCTEROTOME WAS USED DURING A SPHINCTEROTOMY PROCEDURE ON (B)(6), 2011. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE DEVICE FAILED TO BOW COMPLETELY. THE PROCEDURE WAS COMPLETED WITH ANOTHER DREAMTOME RX SPHINCTEROTOME. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. THIS EVENT HAS BEEN DEEMED REPORTABLE BASED ON THE INVESTIGATION RESULTS THAT THE EXTRUSION IS MELTED/SPLIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DREAMTOME RX SPHINCTEROTOME | UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) | KNS | BOSTON SCIENTIFIC - SPENCER | M00584040 | 13818086 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR |