PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2011-02375
- Event Type
- Injury
- Date Received
- April 5, 2011
- Date of Event
- January 10, 2011
- Report Date
- March 11, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.
(B)(4). EVALUATION SUMMARY: EVALUATION OF THE RETURNED COMPONENTS OF THE DEVICE FOUND, THE HANDLE TO FOOT FUNCTION, GUIDE TUBE, NEEDLE GUIDE, BRIDGE, SUTURE BEARING, EXIT RAMP AND SHEATH WERE NORMAL, WITH NO INDICATION OF A PRODUCT QUALITY DEFICIENCY THAT WOULD CONTRIBUTE TO THE REPORTED CUFF MISS. BOTH ENDS OF THE FOOT WERE ALSO EXAMINED AND THERE WERE NO NEEDLE STRIKE MARKS DETECTED. THERE WAS NO MANUFACTURING OR QUALITY INSPECTION DEFICIENCY DETECTED WITH THE RETURNED COMPONENTS THAT WOULD CONTRIBUTE TO THE REPORTED NEEDLE TO CUFF MISS EXPERIENCE. A CUFF-MISS CAN BE INFLUENCED BY MANY FACTORS, INCLUDING, BUT NOT LIMITED TO; MANUFACTURING, FAILURE TO POSITION AND MAINTAIN THE DEVICE AT A 45-DEGREE ANGLE THROUGHOUT DEPLOYMENT, ROTATING THE DEVICE AT ANY POINT DURING PLUNGER/NEEDLE DEPLOYMENT, DEPLOYMENT IN CHALLENGING ANATOMIES (E.G. HEAVY CALCIFIED ARTERIES, MORBIDLY OBESE PATIENTS, ETC.), AGGRESSIVELY DEPLOYING OR REMOVING THE PLUNGER AND / OR INADEQUATE POSITIONING OF THE FOOT AGAINST THE ARTERIAL WALL. TO ASSURE THAT ALL PRODUCTS PERFORM ACCORDING TO SPECIFICATIONS THEY ARE SUBJECT TO AN INSPECTION DURING MANUFACTURING. IN ADDITION, A QUALITY CONTROL AUDIT INSPECTION IS PERFORMED TO VERIFY PRODUCT QUALITY. BASED ON THE INVESTIGATION FINDINGS, THE REPORTED NEEDLE TO CUFF MISS COULD NOT BE CONFIRMED AND NO CAUSE COULD BE DETERMINED. A REVIEW OF THE FINISHED DEVICE LOT HISTORY RECORDS DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. THIS TYPE OF REPORTED ISSUE WILL CONTINUE TO BE MONITORED.
IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE ATTEMPTED PRE-CLOSE SUTURE PLACEMENT FOR ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER AN INTERVENTIONAL VALVULOPLASTY PROCEDURE. REPORTEDLY, A NEEDLE-TO-CUFF MISS OCCURRED. WHEN THE NEEDLE PLUNGER WAS PULLED FOR REMOVAL, NO SUTURE WAS ATTACHED TO THE NEEDLES. THE DEVICE WAS REMOVED OVER A GUIDE WIRE AND A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELA. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 940356H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 97 YR | Required Intervention |