FDA Adverse Event Injury Summary report: N

EQ REV GLENOID PLATE

MDR report key: 20402089 · Received October 8, 2024

Report

Report Number
1038671-2024-03868
Event Type
Injury
Date Received
October 8, 2024
Date of Event
September 30, 2020
Report Date
April 22, 2025
Manufacturer
EXACTECH, INC.
Product Code
KWT
UDI-DI
10885862086488
PMA / PMN Number
K063569
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REVISION REPORTED MAY HAVE BEEN DUE TO GLENOID OR SCREW LOOSENING. HOWEVER, THE REPORTED GLENOID LOOSENING COULD NOT BE CONFIRMED. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES, RADIOGRAPHS, AND RELEVANT CLINICAL INFORMATION WERE NOT PROVIDED.

Additional Manufacturer Narrative · 0

(D10): CONCOMITANT DEVICE(S): 300-01-15 - EQUINOXE, HUMERAL STEM PRIMARY, PRESS FIT 15MM: 5800029 IMPLANT DATE: ON (B)(6) 2019). 320-10-00 - EQUINOXE REVERSE TRAY ADAPTER PLATE TRAY +0: 6152915. 320-15-01 - EQ REV GLENOID PLATE: 6174681. 320-15-05 - EQ REV LOCKING SCREW: 6152475. 320-38-00 - EQUINOXE REVERSE 38MM HUMERAL LINER +0: 6151762.

Description of Event or Problem · 0

APPROXIMATELY 11 MONTH(S) AND 21 DAY(S) POST-OPERATIVE OF A PREVIOUSLY REVISED LEFT RTSA, IT WAS REPORTED VIA CLINICAL STUDY THAT THE PATIENT EXPERIENCED ASEPTIC GLENOID LOOSENING. X-RAYS SHOWED A LOOSE GLENOID SCREW, AND FURTHER IMAGING SHOWED LOOSENING OF THE GLENOID. THE PREVIOUS REVISION WAS DUE TO A SUBSCAPULARIS TEAR OF THE ROTATOR CUFF. THE PATIENT UNDERWENT A STANDARD REVERSE REVISION WITH THE REMOVAL OF THE HUMERAL TRAY, HUMERAL LINER, GLENOSPHERE AND GLENOID BASE PLATE. THE OUTCOME OF THIS EVENT IS CONSIDERED RESOLVED AND THE CLINICAL REPORT INDICATES THAT THIS EVENT IS POSSIBLY RELATED TO THE DEVICE(S) AND/OR TO THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21975 EQ REV GLENOID PLATE PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED KWT EXACTECH, INC. 10885862086488

Patients

Seq Age Sex Outcome Treatment
1 62 YR Male Required Intervention SEE H11