FDA Adverse Event Malfunction Summary report: N

TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM

MDR report key: 2040155 · Received April 5, 2011

Report

Report Number
2134265-2011-01026
Event Type
Malfunction
Date Received
April 5, 2011
Date of Event
March 11, 2011
Report Date
March 11, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE STENT DELIVERY SYSTEM (SDS) WAS SEPARATED IN TWO PIECES AT THE MID-SHAFT WHICH WAS NECKED DOWN/STRETCHED AT THE SEPARATION SITE. THE BALLOON WAS TIGHTLY FOLDED WITH THE STENT IN BETWEEN THE MARKERBANDS. FURTHER INSPECTION PRESENTED NO OTHER DAMAGE OR IRREGULARITIES. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE A SHAFT BREAK OCCURRED. WHILE UNPACKING THE 3.50X28MM TAXUS LIBERTE STENT DELIVERY SYSTEM (SDS) IT WAS NOTED THAT THE SHAFT WAS BROKEN. THE DEVICE NEVER ENTERED THE PATIENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT CONDITION IS OKAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE A SHAFT BREAK OCCURRED. WHILE UNPACKING THE 3.50X28MM TAXUS LIBERTE STENT DELIVERY SYSTEM (SDS) IT WAS NOTED THAT THE SHAFT WAS BROKEN. THE DEVICE NEVER ENTERED THE PATIENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT CONDITION IS OKAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT SYSTEM CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - MAPLE GROVE H7493893628350 12826386

Patients

Seq Age Sex Outcome Treatment
1