PULSAR MAX II
Report
- Report Number
- 2124215-2011-00390
- Event Type
- Injury
- Date Received
- April 5, 2011
- Date of Event
- January 1, 2011
- Report Date
- June 24, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS EVENT WILL BE UPDATED AND RESUBMITTED.
THE PRODUCT HAS NOT BEEN RETURNED. IF RETURNED, OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL UPDATED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT WITH THIS IMPLANTABLE DEVICE EXPERIENCED SYNCOPE. UPON INTERROGATION, EPISODE REVIEW REVEALED LOSS OF CAPTURE. CURRENTLY THIS DEVICE REMAINS IMPLANTED AND IN SERVICE. NO ADDITIONAL PATIENT EFFECTS REPORTED.
INFORMATION WAS LATER RECEIVED THAT THIS DEVICE WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION. NO ADVERSE PATIENT EFFECTS WERE NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PULSAR MAX II | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | 1280 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Life Threatening | MISMATCH| 1280 |