FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2040001 · Received March 25, 2011

Report

Report Number
2531779-2011-01911
Event Type
Malfunction
Date Received
March 25, 2011
Report Date
February 25, 2011
Manufacturer
ANIMAS CORP
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. IT WAS FOUND THAT ALL KEYPAD BUTTONS WERE MISALIGNED AND THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BUTTONS ARE STICKING. IT WAS REPORTED THAT THE KEYPAD IS INTACT AND THERE IS NO EXPOSURE TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1