FDA Adverse Event Injury Summary report: N

0009613348-2024-013451

MDR report key: 20399842 · Received October 8, 2024

Report

Report Number
0009613348-2024-013451
Event Type
Injury
Date Received
October 8, 2024
Date of Event
June 8, 2023
Report Date
October 8, 2024
Manufacturer
INSTITUT STRAUMANN AG
Product Code
DZE
PMA / PMN Number
K150938
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE BATCH NUMBER COULD NOT BE VERIFIED DUE TO INCOMPLETE OR MISSING INFORMATION AND / OR PRODUCT FROM THE CUSTOMER. OUR MANUFACTURING Q-SYSTEM ASSURES THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFIRM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED. THE NON-INTEGRATION OF AN ENDOSSEOUS DENTAL IMPLANT DURING THE HEALING PHASE IS A KNOWN INHERENT RISK OF THE TREATMENT WITH DENTAL IMPLANTS. MOST IMPLANT FAILURES OCCUR BEFORE OCCLUSAL LOADING (ANALAINEN ET AL. 2013, KOLDSLAND ET AL. 2009). BASED ON CLINICAL STUDIES ABOUT 1-3% OF THE IMPLANTS FAIL WITHIN THE FIRST YEAR AFTER IMPLANTATION (GANELES ET AL. 2008). IMPLANTS MAY HAVE TO BE REMOVED IN CASE ONE OR MORE OF THE IMPLANT SUCCESS CRITERIA ARE NOT MET. IMPLANT SUCCESS CRITERIA ACCORDING TO BUSER ET AL. (1991) ARE: 1. ABSENCE OF PERSISTENT SUBJECTIVE COMPLAINTS SUCH AS PAIN, FOREIGN BODY SENSATION AND /OR DYSESTHESIA 2. ABSENCE OF A RECURRENT PERI-IMPLANT INFECTION WITH SUPPURATION 3. ABSENCE OF IMPLANT MOBILITY 4. ABSENCE OF A CONTINUOUS RADIOLUCENCY AROUND THE IMPLANT. THE MANUFACTURER'S TREND ANALYSIS CONFIRMS THAT THE REPORTED EARLY FAILURE RATE ASSOCIATED WITH ITS DENTAL IMPLANTS IS BELOW THE EXPECTED FAILURE RATE FOR THIS TREATMENT AS PUBLISHED IN THE SCIENTIFIC LITERATURE. NON-FATAL SERIOUS INJURY OR DEVICE MALFUNCTION STORED DUE TO COVID 19 PANDEMIC IN ACCORDANCE WITH FDA GUIDANCE "POSTMARKETING ADVERSE EVENT REPORTING FOR MEDICAL PRODUCTS AND DIETARY SUPPLEMENTS DURING A PANDEMIC" PUBLISHED MAY 2020.

Description of Event or Problem · 0

THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 45. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2023 NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: MOBILITY. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
21714 DZE INSTITUT STRAUMANN AG

Patients

Seq Age Sex Outcome Treatment
1 44 YR Male Required Intervention