FDA Adverse Event Malfunction Summary report: N

TRANS-RAY PLUS 7.5FR 40CC IAB WITH ACCESSORIES

MDR report key: 20398516 · Received October 8, 2024

Report

Report Number
2248146-2024-0000617
Event Type
Malfunction
Date Received
October 8, 2024
Date of Event
September 19, 2024
Report Date
December 2, 2024
Manufacturer
DATASCOPE CORP.
Product Code
DSP
UDI-DI
10607567111728
PMA / PMN Number
K122628
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE IAB WAS RETURNED WITH THE MEMBRANE COMPLETELY UNFOLDED AND BLOOD ON THE EXTERIOR AND INTERIOR OF THE CATHETER AND BETWEEN THE CATHETER AND THE SHEATH. THE RETURNED SHEATH WAS OVER THE CATHETER AND COVERING HALF OF THE BALLOON. A KINK WAS OBSERVED ON THE CATHETER TUBING APPROXIMATELY 26.2CM FROM IAB TIP. ADDITIONALLY THE OPTICAL FIBER WAS FOUND TO BE BROKEN WITHIN THE MEMBRANE APPROXIMATELY 5.6CM FROM IAB TIP. THE INNER LUMEN WAS FOUND OCCLUDED WITH DRIED BLOOD. THE OCCLUSION WAS UNABLE TO BE CLEARED. THE OPTICAL FIBER WAS FOUND TO BE BROKEN, CONFIRMING THE REPORTED PROBLEM. WE ARE UNABLE TO DETERMINE WHEN THIS MAY HAVE OCCURRED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT. NO NONCONFORMANCES WERE FOUND THAT ARE CONSIDERED TO BE RELATED TO THE EVENT. REFERENCE COMPLAINT #(B)(4)

Additional Manufacturer Narrative · 0

EVENT SITE POSTAL CODE: 596-0042 THE PRODUCT HAS BEEN RETURNED TO THE MANUFACTURER, BUT IS PENDING INVESTIGATION. ONCE THE INVESTIGATION IS COMPLETED A SUPPLEMENTAL REPORT WITH OUR FINDINGS WILL BE SUBMITTED. COMPLAINT RECORD ID # (B)(4).

Description of Event or Problem · 0

N/A

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING INTRA-AORTIC BALLOON (IAB) THERAPY, A FIBER OPTIC SENSOR FAILURE ALARM OCCURRED. THE WAVEFORM WAS THEN OBTAINED FROM THE ARTERIAL LINE AND THERAPY WAS CONTINUED. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
16666 TRANS-RAY PLUS 7.5FR 40CC IAB WITH ACCESSORIES SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. 0684-00-0608 3000327934 10607567111728

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown