TDCH THERMODILUTION PA CATHETER
Report
- Report Number
- 2025816-2011-00001
- Event Type
- Other
- Date Received
- March 21, 2011
- Date of Event
- November 18, 2010
- Report Date
- December 15, 2010
- Manufacturer
- ICU MEDICAL, INC.
- Product Code
- FOZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
COMPLAINT ANALYSIS: THE THERMISTOR IS A PURCHASED COMPONENT. SUPPLIER QUALITY MGMT. WAS NOTIFIED OF THESE FINDINGS FOR THEIR REVIEW AND FOLLOW UP INFO. CONCLUSION: THE ROOT CAUSE OF THE REPORTED PRODUCT EXPERIENCE CANNOT BE DETERMINED WITH ABSOLUTE CERTAINTY. THE MFR IS MONITORING AND TRENDING FOR ANY REPEAT OCCURRENCES.
UFMW REC'D REPORTING READING ERRORS WITH ONE 41232-01 THERMODILUTION PA CATHETER. IT WAS REPORTED "THE SWAN GANZ CATHETER WAS FINE YESTERDAY AND LAST NIGHT WHEN I HAD THE PT. HOWEVER, THIS MORNING I COULDN'T DO A CARDIAC OUTPUT AND THOUGHT IT WAS THE THERMISTOR. I CHANGED CABLES, THREE DIFFERENT ONES, AND AT ONE POINT DID GET A CARDIAC OUTPUT AND CARDIAC INDEX, BUT THE WAVEFORM NEVER RETURNED TO BASELINE." THE CATHETER WAS PRE-TESTED, INSERTED AND SUCCESSFULLY USED FOR 24 HOURS WITH NO PROBLEMS NOTED. THERE WERE NO REPORTED ADVERSE PT CONSEQUENCES. RECORD REVIEW: A REVIEW OF THE MFG LOT DATABASE FOR LOT # 80-670-LA (MFG DATE 09/2009) SHOWS (B)(4) UNITS WERE MFG. THE 41232-01 CATHETER WAS RETURNED TO THE MFR FOR ANALYSIS AND INVESTIGATION. VISUAL ANALYSIS OF THE RETURNED 41232-01 CATHETER RECORDED NO VISIBLE ANOMALIES. FUNCTIONAL TESTS WERE PERFORMED. THE ENGINEERING REPORT DOCUMENTED INTERMITTENT READINGS WERE REPLICATED WHEN MANIPULATING THE DEVICE THERMISTOR COMPONENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TDCH THERMODILUTION PA CATHETER | PA CATHETER | FOZ | ICU MEDICAL, INC. | 41232-01 | 80-670-LA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | CABLES| UNKNOWN: MANUFACTURER/MAKE/MODEL| UNKNOWN: CARDIAC OUTPUT COMPUTER |