IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM
Report
- Report Number
- 0002023141-2024-03197
- Event Type
- Injury
- Date Received
- October 8, 2024
- Date of Event
- September 11, 2024
- Report Date
- February 17, 2025
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024019508
- PMA / PMN Number
- K061410
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED. G4: PREMARKET IDENTIFICATION K011028/K013227.
ZIMVIE COMPLAINT NUMBER (B)(4). ZIMVIE RECEIVED ONE (1) TSVB10, (IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM), FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED, APPARENT BONE / TISSUE ATTACHED TO EXTERNAL THREADS, AND WAS OBSERVED FRACTURED AT THE COLLAR REGION. THERE WAS A FRACTURED SCREW STUCK INSIDE THE IMPLANT. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1253501. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1253501, FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : FUNCTIONAL : FRACTURE : IMPLANT BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS CLINICIAN SELECTS IMPLANT THAT IS UNABLE TO RESIST LONG-TERM OCCLUSAL LOADING. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE FRACTURE HAS BEEN IDENTIFIED ON THE COLLAR. THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.
DOCTOR REPORTED IMPLANT COLLAR FRACTURE, TOOTH # 45 AND WAS REMOVED.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 217273 | IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | 1253501 | 00889024019508 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Male |