FDA Adverse Event Injury Summary report: N

IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM

MDR report key: 20394169 · Received October 8, 2024

Report

Report Number
0002023141-2024-03197
Event Type
Injury
Date Received
October 8, 2024
Date of Event
September 11, 2024
Report Date
February 17, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019508
PMA / PMN Number
K061410
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). A4: PATIENT WEIGHT UNKNOWN / NOT PROVIDED. G4: PREMARKET IDENTIFICATION K011028/K013227.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). ZIMVIE RECEIVED ONE (1) TSVB10, (IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM), FOR EVALUATION. VISUAL EVALUATION WAS PERFORMED, APPARENT BONE / TISSUE ATTACHED TO EXTERNAL THREADS, AND WAS OBSERVED FRACTURED AT THE COLLAR REGION. THERE WAS A FRACTURED SCREW STUCK INSIDE THE IMPLANT. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1253501. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1253501, FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : FUNCTIONAL : FRACTURE : IMPLANT BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS CLINICIAN SELECTS IMPLANT THAT IS UNABLE TO RESIST LONG-TERM OCCLUSAL LOADING. THEREFORE, BASED ON THE AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE FRACTURE HAS BEEN IDENTIFIED ON THE COLLAR. THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

DOCTOR REPORTED IMPLANT COLLAR FRACTURE, TOOTH # 45 AND WAS REMOVED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
217273 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM DENTAL IMPLANT DZE ZIMMER DENTAL 1253501 00889024019508

Patients

Seq Age Sex Outcome Treatment
1 62 YR Male