FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2039255 · Received March 25, 2011

Report

Report Number
3004209178-2011-80819
Event Type
Injury
Date Received
March 25, 2011
Date of Event
March 11, 2011
Report Date
March 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 560 MG/DL. THE CUSTOMER STATED THAT PRIOR TO THE EVENT, SHE THOUGHT SHE HAD STOMACH FLU. THE CUSTOMER ALSO STATED THAT SHE HAS BEEN GETTING EXCESSIVE NO DELIVERY ALARMS. THE CUSTOMER FURTHER STATED THAT SHE USES AN MMT-398 QUICK-SET INFUSION SET, LOT 9201593. TROUBLESHOOTING WAS PERFORMED. THE INSULIN PUMP PASSED BOTH THE FIXED PRIME AND THE HIGH PRESSURE TEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 21 YR Hospitalization UNOMEDICAL PARADIGM QUICK-SET| INSULIN INFUSION SET, LOT #9201593