30K FSI-PWR-FG-10 INSERT,PKD
Report
- Report Number
- 2424472-2024-00398
- Event Type
- Malfunction
- Date Received
- October 7, 2024
- Report Date
- December 28, 2024
- Manufacturer
- DENTSPLY LLC
- Product Code
- ELC
- UDI-DI
- D003820021
- PMA / PMN Number
- K052334
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
RETURN QTY. 1, 23270, 30K FS-PWR-10 FG-27, 00107007 BUL. 30K. TEST METHOD / GP005-WI02. INDUCTANCE: GAUGE ID#: 5349, DUE: 11/30/2024, RESULT 3.14. MEETS SPEC, DOES NOT MEET SPEC, N/A. TIP CONDITION: MEETS SPEC, DOES NOT MEET SPEC, N/A. STACK CONDITION, MEETS SPEC, DOES NOT MEET SPEC, N/A. TESTED ON: G130 GAGE ID#: 9626-2, DUE DATE: 03/31/2025, SET PRESSURE: 35 PSI. WATER FLOW: OUTPUT RATE: 35, PSI, MEETS SPEC, DOES NOT MEET SPEC, N/A. SPRAY: MEETS SPEC, DOES NOT MEET SPEC, N/A. WATER LEAK: HP SEALING RING, PROXIMAL RING, DISTAL RING. SEAM LEAK, N/A. VIBRATION: MEETS SPEC, DOES NOT MEET SPEC, N/A. GRIP CONDITION: MEETS SPEC, DOES NOT MEET SPEC, N/A. OTHER VISUAL OBSERVATION: INSERT TIP IS CLOGGED, NO WATER FLOW. ALLEGED EVENT: CONFIRMED, NOT CONFIRMED.
IN THIS EVENT IT IS REPORTED THAT A 30K FSI-PWR-FG-10 INSERT,PKD HAS NO WATER FLOW AND GETS HOT. NO INJURY REPORTED SINCE IT WAS NOT USED ON A PATIENT.