KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Report
- Report Number
- 1627487-2024-11367
- Event Type
- Injury
- Date Received
- October 7, 2024
- Date of Event
- September 17, 2024
- Report Date
- October 7, 2024
- Manufacturer
- ABBOTT MEDICAL
- Product Code
- PMP
- UDI-DI
- 05415067027153
- PMA / PMN Number
- P150004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
B3-DATE OF EVENT IS ESTIMATED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT: COMMON DEVICE NAME: DRG: LEAD, MODEL: MN20450-50A, UDI: (B)(4), SERIAL: (B)(6), BATCH: 7982307. COMMON DEVICE NAME: DRG: LEAD, MODEL: MN20450-50A, UDI: (B)(4), SERIAL: (B)(6), BATCH: 7982304. A PATIENT EXPERIENCING HIGH IMPEDANCE WAS REPORTED TO ABBOTT. THE PATIENT LEADS WERE EXPLANTED AND REPLACED. THERAPY WAS RESTORED. THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICES WERE NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED.
IT WAS REPORTED THAT PATIENT EXPERIENCED INEFFECTIVE THERAPY. SYSTEM DIAGNOSTICS RECORDED HIGH IMPEDANCES ON THE LEFT T12 LEAD AND RIGHT S1 LEAD. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEADS WERE EXPLANTED AND REPLACED. EFFECTIVE THERAPY WAS RESTORED POST OPERATIVELY. THE INVESTIGATION WAS UNABLE TO DETERMINE THE LEADS THAT WERE ASSOCIATED WITH THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 217163 | KIT IMPLANTABLE SLIM TIP LEAD, 50CM | DRG LEAD | PMP | ABBOTT MEDICAL | MN10450-50A | 7982304 | 05415067027153 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Male | Other | DGR IPG| DRG LEAD X2 |