PUMP 1886 780G OUS BLE PUMP MG/DL
Report
- Report Number
- 2032227-2024-249653
- Event Type
- Malfunction
- Date Received
- October 7, 2024
- Date of Event
- September 16, 2024
- Report Date
- December 9, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
UNIT PASSED ALL FOLLOWING TESTS: DISPLACEMENT, REWIND, PRIME/SEATING, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, AND SELF TEST. NO UNEXPECTED INSULIN FLOW BLOCK ALARMS NOTED DURING TESTING. UNIT WAS SUCCESSFULLY DOWNLOADED USING THUMP SOFTWARE. HOWEVER ON ADAPT, CONFIRMED NO DELIVERY (7) ALARM ON 09/16/2024 14:05:35.000, 09/16/2024 14:07:00.000, 09/16/2024 14:07:32.000, AND 09/16/2024 14:11:07.000 ALL DURING BOLUS. PUMP WAS CUT OPEN TO PERFORM A VISUAL INSPECTION AND FOUND NO MOISTURE OR PHYSICAL DAMAGE. P-CAP LOCKED IN PLACE PROPERLY DURING TESTING. THE FOLLOWING COSMETIC DAMAGES WERE NOTED: CRACKED KEYPAD OVERLAY AND SCRATCHED CASE. IN SUMMARY, CUSTOMER CONCERN FOR NO DELIVERY/OCCLUSION ALARM IS NOT CONFIRMED. UNIT FUNCTIONED PROPERLY AND PASSED ALL TESTING WITHIN SPEC. NO ALARMS NOTED DURING TESTING, HOWEVER NO DELIVERY (7) NOTED IN DOWNLOAD HISTORY REVIEW ON EVENT DATE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED INSULIN FLOW BLOCK ALARM. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1886. TROUBLESHOOTING WAS PERFORMED AND CUSTOMER CONFIRMED INSULIN DID NOT EXIT DURING 5U FILL CANNULA OR PUMP ALARMED. CUSTOMER RE-ATTEMPTED LOAD RESERVOIR/FILL TUBING PROCESS AFTER A COMPLETE SET CHANGE AND INSULIN DID NOT EXIT OR PUMP ALARMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1886 WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 213991 | PUMP 1886 780G OUS BLE PUMP MG/DL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1886 | HG526V9ZZ |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR | Unknown |