FDA Adverse Event Malfunction Summary report: N

I-STAT EC8+ CARTRIDGE

MDR report key: 20389327 · Received October 7, 2024

Report

Report Number
2245578-2024-00186
Event Type
Malfunction
Date Received
October 7, 2024
Date of Event
August 31, 2024
Report Date
November 22, 2024
Manufacturer
ABBOTT POINT OF CARE
Product Code
CHL
PMA / PMN Number
K940918
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 21-NOV-2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV AN, PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED FOR EC8+ CARTRIDGE LOT K24082 AND CG8+ CARTRIDGE LOT W24205.

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

NA.

Description of Event or Problem · 0

ON 01-OCT-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EC8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT POTASSIUM RESULTS ON A 50 YEAR OLD MALE PATIENT WITH PNEUMONIA, AND A TRAUMA PATIENT.. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: I-STAT, DATE: (B)(6) 2024, COLLECTED: 20:26; 20:27, TESTED: 7.39, PH: A, SAMPLE: HEMOLYZED. METHOD: I-STAT, DATE: (B)(6) 2024, COLLECTED: 20:26; 20:35, TESTED: 7.33, PH: A. METHOD: I-STAT, DATE: (B)(6) 2024, COLLECTED: 20:50; 20:52, TESTED: 7.144, PH: B. METHOD: GEM 5000, DATE: (B)(6) 2024, COLLECTED: 20:50; 20:58, TESTED: 7.31, PH: B. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2546628 I-STAT EC8+ CARTRIDGE EC8+ CARTRIDGE CHL ABBOTT POINT OF CARE NA K24082

Patients

Seq Age Sex Outcome Treatment
1 50 YR Male